V-760 UV-Visible/NIR Spectrophotometer

Features
System Description
The optical bench in the V-760 UV-visible spectrophotometer utilizes a double monochromator to provide exceptional resolution with extremely low stray light (0.00008%) to assure more precise measurement over the widest photometric range. The V-760 provides excellent sensitivity for diffuse reflectance or transmittance of solid and liquid samples using an integrating sphere. The spectral bandwidth can be set as narrow as 0.1 nm for high resolution measurements, such as gas and vapor phase spectroscopy. Extended photometric linearity up to 8 Abs enables measurement of highly absorbing samples and is also effective for measurements of optical material with transmission characteristics that approach 0 %T.
Operating Systems
Two user interfaces are available. The Spectra Manager™ II software, the latest version of JASCO's innovative cross-platform spectroscopy software, is now compatible with Windows 7 Pro (32- and 64-bit configurations) and the Windows 8.1 operating systems.
For simple and routine operation, the intelligent remote module (iRM) has been completely redesigned with a new look and feel, with a color LCD touch screen. Data can also be downloaded to a USB memory stick for portability or further PC data processing.
The V-700 Series has a growing list of software applications for both Spectra Manager II and for the iRM. If you have an application which you don’t see listed, please let us know as we may already have it or we can prepare an application designed specifically for your requirements.
Both of these interfaces allow full system control and advanced data processing. Spectra Manager CFR is the 21 CFR part 11 compliant version of software and is available as an option.
System Features
IQ Accessory and IQ Start
The IQ Accessory function provides automated accessory recognition with identification of the accessory within the method. The spectral data files include the accessory information and serial number while designated instrument parameters can be configured and loaded for optimum sample measurements. IQ Start together with the Start Button is used to immediately start a sample method for immediate access when making routine measurements.
Excellent Optical Performance
The most recent advances in optical technology have been utilized to provide excellent instrument reliability and to assure highly accurate results. High throughput optics combined with a modern electronic design provide excellent sensitivity and stability for all manner of sample measurements.
Full Line of Sampling Accessories
A full complement of accessories is available for the V-700 series of UV-Visible-NIR spectrophotometers for virtually any biological, material science or routine QA/QC measurement. Some accessories are common across the V-700 Series and others are instrument specific. Refer to the instrument brochure for accessories specific to your desired instrument.
Options include a wide variety of liquid cell holders, micro and ultra-micro cell holders, flow cell units and accessories for solid samples. Advanced accessories, include automated cell changers, sippers, integrating spheres, automated and manual absolute reflectance accessories and programmable temperature control systems are available for any of the V-700 instrument systems.
V-760 UV-Vis Spectrophotometer Specifications
Optical System | Czerny-Turner mount Double monochromator Fully symmetrical double beam type |
Light Source | Halogen lamp, Deuterium lamp |
Wavelength range | 187 to 900 nm |
Wavelength accuracy | +/-0.1 nm (at 656.1 nm) |
Wavelength repeatability | +/-0.05 nm |
Spectral bandwidth (SBW) | 0.1, 0.2, 0.5, 1, 5, 2, 10 nm L2, L5, L10 nm (low stray light mode) M1, M2 nm (micro cell mode) |
Stray light | 1 % (198 nm KCL 12 g/L aqueous solution) 0.00008 % (220 nm NaI 10 g/L aqueous solution) 0.00008 % (340 nm NaNO2 50 g/L aqueous solution) 0.00008 % (370 nm NaNO2 50 g/L aqueous solution) SBW: L2 nm |
Photometric range | -4~6 Abs |
Photometric accuracy | +/-0.0015 Abs (0 to 0.5 Abs) +/-0.0025 Abs (0.5 to 1 Abs) +/-0.3 %T Tested with NIST SRM 930D |
Photometric repeatability | +/-0.0005 Abs (0 to 0.5 Abs) +/-0.0005 Abs (0.5 to 1 Abs) Tested with NIST SRM 930D |
Scanning speed | 10~4000 nm/min (8000 nm/min in preview mode) |
Slew speed | 12,000 nm/min |
RMS noise | 0.00003 Abs (0 Abs, wavelength: 500 nm, measurement time: 60 sec, SBW: 2 nm) |
Baseline stability | 0.0003 Abs/hour (Value obtained more than two hours after turning on the source, when the room temperature is stabilized, wavelength: 250 nm, response: slow and SBW: 2 nm) |
Baseline flatness | +/-0.0003 Abs (200 - 800 nm) |
Detector | Photomultiplier tube |
Standard facilities | |
Standard program | |
Dimensions and Weight | 460(W) x 602(D) x 268(H) mm, 29 kg |
Power requirements | 150 VA |
Installation requirements | Room temperature: 15-30 Celsius, humidity: below 85% |
Applications
Click one of the UV-VIS applications below for more information.
Automated Sun Protection Factor (SPF) Analysis of Sunscreen using UV-Vis
Measurement of Enzyme Activity - ALP Activity
Quantitative Determination of Proteins
TCH-703 8-position Turret Micro Cell Holder
Measurement reproducibility using SAH-769 one-drop measurement unit
The Model SAH-769 One-Drop: A Dedicated Accessory for Extremely Small Amounts of Protein and Nucleic Acid Samples
Australia
New Zealand
Australia Freephone
1800 00 84 53
sales.au@bio-strategy.com
New Zealand Freephone
0800 34 24 66
sales.nz@bio-strategy.com
Bio-Strategy PTY Limited
MELBOURNE (HQ)
8 – 10 Trade Park Drive
Tullamarine, VIC 3043
Australia
BRISBANE
2/7 Miller Street
Murarrie, QLD 4172
Australia
Bio-Strategy Limited
33 Westpoint Drive
Hobsonville
Auckland 0618
New Zealand
Advanced Analytical Solutions For Your Science

Copyright © 2023 Bio-Strategy Limited. Terms & Conditions