Skip to content
Bio-Strategy Australia 1800 00 84 53 | Bio-Strategy New Zealand 0800 34 24 66

V-1000 Vacuum Ultraviolet Spectrophotometer

semiconductor-vacuums-1

The V-1000 provides rapid sample measurements without evacuation of the entire instrument by using a nitrogen purge and dual sample exchange cabinets to enhance purge efficiency.

Features

System Description

Miniaturization of wafer manufacturing processes has promoted the shift towards shorter photolithography wavelengths. Currently, ArF lasers (193 nm) are most currently used. Research on the practical implementation of F2 (157 nm) has reduced the imprinting node to 0.07 µm.

The V-1000 was designed to evaluate optical properties in the wavelength range between 115 and 300 nm. Typically, instrument optical paths for analyses in this range are evacuated due to the strong absorption of oxygen, thus, the ‘vacuum’ ultraviolet range. Superior photometric reproducibility is achieved by using a double monochromator for elimination of stray light and a reference stabilized double-beam system to maintain baseline stability.

System Features

  • Highly Accurate Measurements in the vacuum UV range

Double monochromator and reference stabilized double-beam system achieves a photometric reproducibility of 0.03% .

  • Rapid switchover between transmission/reflection measurement
 

Samples are easily loaded for either reflection measurements at 5° or 45° or transmission measurements without replacing sampling accessories. Please contact us if you require a special configuration.

  • Quick measurements
 

By increasing purge efficiency and eliminating dead space in the optical path, measurements are completed within minutes.

V-1000 Specifications

Wavelength Range 115 to 300 nm
Photometric Reproducibility Standard deviation at 157 nm: ±0.03 % or less for material with 70 %T or greater
Spectrophotometer Double monochromator with dual concave gratings and reference stabilized double-beam system
Instrument Resolution 3 nm (Optional: 1 nm)
Light Source Deuterium lamp
Detector Photomultiplier tube (PMT)
Purge Nitrogen purge (Sample chamber purged separately)
Measurement Configuration Transmission and 5° or 45° incidence reflection (standard)
Optional Accessories Absolute reflection and polarization dependence
Sample Size Maximum sample size 50.8 mm diameter and 10 mm thickness
Data Collection and Analysis Software Spectral scanning, fixed wavelength, baseline correction, smoothing, spectral arithmetic, peak area, peak height, peak half width, and data conversion routines
Supplied Support Equipment Support bench, oxygen monitor, vacuum gauge and vacuum pump
Size 1300(W) x 700(D) x 1200(H)
Weight Approx. 450 kg
Power Requirement 130 VA

Details

Superb Photometric Reproducibility

This example uses a calcium fluoride window 30 mm in diameter and 2 mm thick.
One hundred consecutive measurements were taken on five separate occasions using a wavelength of 157.6 nm.

Measurement
Average Transmittance
Standard Deviation
(Num. of samples: 100)
1
86.07 %
0.025 %
2
86.03 %
0.023 %
3
86.02 %
0.023 %
4
86.04 %
0.024 %
5
86.05 %
0.025 %

%T Transmission measurement

transmission-measurement

%R Reflection measurement

reflection-measurement

Australia

Learn more ›

New Zealand

Learn more ›

Australia Freephone

1800 00 84 53

sales.au@bio-strategy.com

 

New Zealand Freephone

0800 34 24 66

sales.nz@bio-strategy.com

 

Bio-Strategy PTY Limited

MELBOURNE (HQ)

8 – 10 Trade Park Drive

Tullamarine, VIC 3043

Australia

 

BRISBANE

2/7 Miller Street

Murarrie, QLD 4172

Australia

 

Bio-Strategy Limited

33 Westpoint Drive

Hobsonville

Auckland 0618

New Zealand

Advanced Analytical Solutions For Your Science

Electromag-01

Copyright © 2023 Bio-Strategy Limited. Terms & Conditions

Scroll To Top