V-1000 Vacuum Ultraviolet Spectrophotometer

The V-1000 provides rapid sample measurements without evacuation of the entire instrument by using a nitrogen purge and dual sample exchange cabinets to enhance purge efficiency.
Features
System Description
Miniaturization of wafer manufacturing processes has promoted the shift towards shorter photolithography wavelengths. Currently, ArF lasers (193 nm) are most currently used. Research on the practical implementation of F2 (157 nm) has reduced the imprinting node to 0.07 µm.
The V-1000 was designed to evaluate optical properties in the wavelength range between 115 and 300 nm. Typically, instrument optical paths for analyses in this range are evacuated due to the strong absorption of oxygen, thus, the ‘vacuum’ ultraviolet range. Superior photometric reproducibility is achieved by using a double monochromator for elimination of stray light and a reference stabilized double-beam system to maintain baseline stability.
System Features
- Highly Accurate Measurements in the vacuum UV range
Double monochromator and reference stabilized double-beam system achieves a photometric reproducibility of 0.03% .
- Rapid switchover between transmission/reflection measurement
Samples are easily loaded for either reflection measurements at 5° or 45° or transmission measurements without replacing sampling accessories. Please contact us if you require a special configuration.
- Quick measurements
By increasing purge efficiency and eliminating dead space in the optical path, measurements are completed within minutes.
V-1000 Specifications
Wavelength Range | 115 to 300 nm |
Photometric Reproducibility | Standard deviation at 157 nm: ±0.03 % or less for material with 70 %T or greater |
Spectrophotometer | Double monochromator with dual concave gratings and reference stabilized double-beam system |
Instrument Resolution | 3 nm (Optional: 1 nm) |
Light Source | Deuterium lamp |
Detector | Photomultiplier tube (PMT) |
Purge | Nitrogen purge (Sample chamber purged separately) |
Measurement Configuration | Transmission and 5° or 45° incidence reflection (standard) |
Optional Accessories | Absolute reflection and polarization dependence |
Sample Size | Maximum sample size 50.8 mm diameter and 10 mm thickness |
Data Collection and Analysis Software | Spectral scanning, fixed wavelength, baseline correction, smoothing, spectral arithmetic, peak area, peak height, peak half width, and data conversion routines |
Supplied Support Equipment | Support bench, oxygen monitor, vacuum gauge and vacuum pump |
Size | 1300(W) x 700(D) x 1200(H) |
Weight | Approx. 450 kg |
Power Requirement | 130 VA |
Details
Superb Photometric Reproducibility
This example uses a calcium fluoride window 30 mm in diameter and 2 mm thick.
One hundred consecutive measurements were taken on five separate occasions using a wavelength of 157.6 nm.
(Num. of samples: 100) |
||
%T Transmission measurement
%R Reflection measurement
Australia
New Zealand
Australia Freephone
1800 00 84 53
sales.au@bio-strategy.com
New Zealand Freephone
0800 34 24 66
sales.nz@bio-strategy.com
Bio-Strategy PTY Limited
MELBOURNE (HQ)
8 – 10 Trade Park Drive
Tullamarine, VIC 3043
Australia
BRISBANE
2/7 Miller Street
Murarrie, QLD 4172
Australia
Bio-Strategy Limited
33 Westpoint Drive
Hobsonville
Auckland 0618
New Zealand
Advanced Analytical Solutions For Your Science

Copyright © 2023 Bio-Strategy Limited. Terms & Conditions